Product Description
The Serinus 10 Ozone (O3) analyser delivers precise and reliable performance at excellent value. It uses proven non-dispersive ultraviolet (UV) absorption technology to measure O3 in ambient air (LDL < 0.5 ppb, range 0 – 20 ppm).
APPROVALS
• US EPA approval (EQOA–0809–187)
• EN approval TÜV (0000040202-02)
• EN approval MCERTS (MC100165/07)
• Australian Standard (AS 3580.6.1-2011)
• Russian approval (56053-13)
• French approval LSCQA
• Ukraine approval (12/3/B/24/295-17)
• Korean approval NIER (AACMS-2017-03)
BENEFITS
The Serinus® range of analysers have been designed using our
experience and knowledge gained from operating large air quality
monitoring networks for more than 40 years. The result, instruments
that integrate seamlessly into continuous monitoring networks.
• Reliable performance complemented with a two year warranty
• Minimise time spent on site performing maintenance through
superior remote instrument control, diagnostic viewing and
calibration
• Comprehensive data logging and remote viewing of over 200
operational parameters
• Versatile interfacing through RS232, USB, Bluetooth, analog and
digital I/O and optional TCP/IP
• Easy setup through an intuitive menu system, advanced GUI and a
large alphanumeric keypad with tactile keys
• Instant status indication via illuminated instrument status light
buttons on the front panel
• Removable flash memory stores ~10 years of data, including up to
12 individual parameters and event logs. Operational parameters
can also be transferred to memory for easy retrieval
• Latest firmware updates can easily be installed using the USB
flash memory drive
• Programming, viewing, downloading and emailing of data is
made even simpler using Bluetooth connectivity combined with
the ‘Ecotech Serinus® Remote’ Android App
• Built-in digital display measuring values and having function of
measuring values transferred to graphical trend display
• Enhanced operator safety through the use of 12 VDC internal
voltages
• Reduced cost of spare parts, accessories and consumable items
through extensive use of standard components across the
Serinus® range
• Rack slide design makes accessing internal components and
removing the analyser from a rack cabinet easy
• AQMS power usage can be reduced due to the instrument’s low
power demand and its ability to operate over a wide temperature
range.
DATA LOGGING
• 8 GB removable USB flash memory drive that stores the internal
data logger, event log, automatic/manual configuration backup
and automatic/manual parameter list save
• Internal data logger (can log up to 12 of 200+ parameters)
• Logger interval (1 second to 24 hours)
• Storage capacity ~ 10 years (set to 1 minute interval).
OPTIONS & ACCESSORIES
• TCP/IP Network Port
• Dual Sample (particulate) Filter
• High Pressure Zero (for direct cylinder calibration)
• Internal Zero and Span (for instrument precision checks)
• Internal DC Pump (long life, low power, minimal maintenance)
• Metric Fitting Kit (to connect 6 mm tube to rear ports)
• 19” Rack Mount Kit
• External Pump (240 or 110 VAC)
• Annual Maintenance Kit
• Printed User Manual (a soft copy of the user manual is supplied
on the ACOEM Ecotech resources USB stick with each analyser).
COMMUNICATION
• #1 RS232 port (supports all protocols*)
• #2 RS232 port (supports all protocols* and multidrop)
• USB port (Advanced protocol only)
• Bluetooth (Advanced protocol only)
• TCP/IP network port (optional, supports all protocols*).
* Available protocols: Modbus (RTU/TCP), Bayern Hessen, EC9800 or Advanced
COMMUNICATION ANALOG & DIGITAL I/O PORT
• 3 x analog outputs (current or voltage), 12 bit resolution user
definable range (between two user specified points)
Current range: 0 - 20 mA, 2 - 20 mA or 4 - 20 mA
Voltage range: 0 - 5 V, 0 - 10 V (voltage offset of 0 V, 0.25 V or 0.5 V)
• 3 x analog inputs (scalable), 0 - 5 V, 16 bit resolution
• 8 x logic level digital control inputs
• 8 x open collector digital status outputs.