Product Description
Ambient air monitoring is the systematic, long-term assessment of pollutant levels by measuring the quantity and types of certain pollutants in the surrounding, outdoor air. We are the leading supplier of Ambient Air Monitoring Equipment. Instrumex make IPM-FDS(2. 5µ / 10µ) Ambient Fine Dust Dampler
Model- Microprocessor based
PM 2. 5 size separator
EPA PM 2. 5 WINS impactor
PM 10 size separator
EPA omni- directional ambient particle inlet with 10µ separation assembly & sample transport tube.
Filter Media
47 mm filter in a filter cassette housed in filter holder assembly.
Flow rate meter & Volume totaliser
(USEPA procedure followed)
Flow in LPM on digital display with a resolution of 0. 01 LPM under actual operating conditions. Flow rate maintained constant @ 16. 67 LPM with ±2% accuracy throughout sampling period using microprocessor based digital flow controller.
Clock / Timer system
Programmable real time control system with automatic start & stop & digital display of date & time & time of sampling. Accuracy ± 2 min / month.
Temperature sensors
For ambient temperature & Filter temperature range - 5°c to 50°c with a resolution of 0. 10c.
Bar press sensor
Range 600 to 800 mm Hg. Resolution, 1 mm HG
Display
20x4 LCD display showing parameters in real time viz.
· Flow rate (16. 67) LPM
· Volumetric flow
· Ambient temperature
· Filter temperature
· Barometric pressure
Outputterminal
RS-232 port for transferring data into pc.
option for USB available.
Calibrated from
Fluid Control Research Institute, Palakkad, Kerala
Additional Information:
Pay Mode Terms: Other